Leica VT1200 and VT1200S – Microtome
5. Working with the Instrument
Key/setting dial semiautomatic sectioning mode
In semiautomatic sectioning mode,
before each cut, a manual feed to
the desired section thickness must
be carried out using the setting dial
for section thickness.
There is no automatic specimen
retraction in this mode; however,
retraction can be performed man-
ually.
The following key functions are dis-
abled in semiautomatic mode:
Setting cutting window edges
Selecting the continuous stroke
(CONT)
Selecting the section thickness
for the automatic feed (AUTO
FEED)
PAUSE
An audible warning signal sounds
when these keys are pressed.
automatic sectioning mode
In automatic mode, the selected
section thickness feed (AUTO
FEED) is carried out along the 1st
selected cutting window edge
automatically before each cut. To
prevent the specimen surface and
the blade from coming into contact
while the blade is being retracted,
the specimen is lowered by the
desired retraction value along the
2nd cutting window edge after
each complete cut.
The following functions keys be-
come enabled again in automatic
mode:
Cutting window edges that have
already been set
Section thickness (AUTO FEED)
and continuous stroke (CONT)
Toggling from AUTO
to MAN
Toggling from MAN to
AUTO