6. Startup
Insert the centering key through the corre-
sponding openings (44.1) in the condenser
holder.
Turn the centering screws until the dark ring
(phase ring in the objective) is congruent with
the slightly narrower bright ring (light ring in
condenser) (43 c).
Repeat the process for all other phase con-
trast objectives.
Remove the centering keys after the centering
procedure.
Note:
During change of objectives the centering keys
must not remain in the openings of the
condenser.
6.7 Adjusting the Light Sources
Transmitted Light Axis (TL) with 107/2 Lamp
housing
Fig. 44 Light ring centering
1 Clamping screw
1