Leica DM4000M Microscope & Magnifier User Manual


 
39
6. Startup
If your microscope is equipped for the use of
phase contrast, the light rings that fit the objec-
tives are built into the condenser.
The light rings are already leveled in the factory.
However, the leveling should be rechecked.
Note:
Every objective is assigned its own light ring in
the condenser disc. Therefore, a check must be
performed for each objective. When swiveling in
a suitable objective for phase contrast, the cor-
responding light ring is set automatically.
Press the BF (Bright Field) button (one of the
variable function keys, behind the focus dials).
PH LR
Fig. 43 Phase contrast centering procedure
PH=phase contrast ring, LR=light ring
a Condenser in bright field (BF) position
b Condenser in phase contrast (PH) position
Light ring (LR) not centered
c Light ring and phase ring centered
In the place of an eyepiece, insert the focus-
ing telescope (Fig. 42) into the observation
tube.
Swivel in the phase contrast objective with
the least magnification.
Focus the ring structure (43a) by slightly loos-
ening the clamping ring (42.2) and moving the
eyelens (42.1).
Retighten the clamping ring.
Press the PH (Phase Contrast) button. The ring
diaphragm in the condenser is pivoted in.
If the light ring and the phase ring are not
shown as arranged in Fig. 43c, the light ring
must be leveled.
Fig. 42 Focusing telescope
1 Adjustable eyelens
2 Clamping ring for fixing the focus position
AB C
1
2