Invacare TPO100B Oxygen Equipment User Manual


 
SECTION 8—TROUBLESHOOTING GUIDE
Part No 1156872 43 Invacare SOLO
2
LCD DISPLAY: INDICATORS:
Continuous audible beep every half
second.
RED indicator is On.
GREEN indicator is Off.
DESCRIPTION: SOLUTIONS:
Abnormal operating condition has been
detected
1. Turn unit Off, then On again to retry.
2. Change to another source of oxygen if
alarm continues.
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.
Call your provider and report the problem if alarm continues.
LCD DISPLAY: INDICATORS:
Ten audible beeps every 10 sec
RED indicator Flashing Fast.
DESCRIPTION: SOLUTIONS:
Battery pack is too hot, or too cold, to
allow battery to continue to operate.
1. Remove battery and only use AC, or
DC, adapter for power. Or, use an
alternate battery pack.
2. Move battery to warmer, or cooler,
surroundings. Allow battery to cool down
to less than 95° F (35° C), or warm up to
50° F (10° C).
3. Turn unit off. Re-insert battery and
retry.
4. Change to another source of oxygen if
no other power source is available.
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens