SECTION 6—OPERATING INSTRUCTIONS
Invacare SOLO
2
18 Part No 1156872
SECTION 6—OPERATING
INSTRUCTIONS
Location
ƽ WARNING
NEVERblocktheairopeningoftheproductorplaceitona
softsurface,suchas abedorcouch,wherethe airopeningmay
beblocked.Keeptheopeningsfree fromlint,hairandthelike.
Keepunitatleast3in(7.6cm)awayfromwalls,draperies,
furniture,andthelike.
Locateandpositionthetransporta bl eoxygenconcentratorinawell
ventilatedspaceso thattheairintakeandtheairexhaustsarenot
obstructed.
Recommendations for Optimal Performance
Filters
Airentersthetransportableoxygenconcentratorthroughanairintakefilter
locatedontheleftsideoftheunit.Thisf ilterpreventshairandotherlarge
particlesintheairfromenteringtheunit.Beforeyouoperatethe
transportableoxygenconce n tr ator,makesurethefilterisclean,dryand
properlypositioned.
Toclean/replacetheairintakefilter,refertoCleaning, Care,And
Maintenanceonpage 32.
Operating Temperature: 41°F to 95°F (5°C to 35°C)
Relative Humidity: 15% to 60%
Transport/Storage Temperature: -2°F to 140°F (-20°C to 60°C)
Allow unit to warm, or cool, to operating
temperature range before using.
Electrical: No extension cords.
Altitude: Up to 10,000 ft (3046 m) above sea level.
Tubing and Cannula: 7 ft (2.1 m) crush resistant cannula
(DO NOT pinch).
Environment: Smoke, pollutant and fume free. No confined
spaces (example: no closets).
Time of Operation: Up to 24 hours per day when connected to AC or
DC power.