Invacare TPO100B Oxygen Equipment User Manual


 
SECTION 5—TYPICAL PRODUCT PARAMETERS
Invacare SOLO
2
16 Part No 1156872
MAXIMUM RECOMMENDED
FLOWRATE WITH 7KPA (1.01
PSI) BACKPRESSURE:
3.0 LPM
PRESSURE RELIEF
ACTIVATION:
20 psi ± 5 psi (137.8 kPa ± 34.5 kPa)
MAX OUTLET PRESSURE @
3.0 LPM CONTINUOUS:
12.0 psig
DIMENSIONS:
(WITHOUT CART)
16.5 in high x 11 in wide x 8 in deep (41.9 cm
high x 27.9 cm wide x 20.3 cm deep)
WEIGHT:
(UNIT WITHOUT CART)
< 20 lbs (9.09 kg)
BATTERY DURATION:
(TIMES ARE APPROXIMATE)
Pulse Mode Continuous Mode
Setting 1 = 4.5 hrs Setting 0.5 LPM = 4.5
hrs
Setting 2 = 3.5 hrs Setting 1 LPM = 3.5 hrs
Setting 3 = 3.0 hrs Setting 2 LPM = 2.5 hrs
Setting 4 = 2.5 hrs Setting 3 LPM = 1.5 hr
Setting 5 = 2.5 hrs
BATTERY RECHARGE TIME: 5 hours
Recharge time increases if battery is charging
while unit is running.
HUMIDITY RANGE:
OPERATING HUMIDITY:
TRANSPORT AND STORAGE:
15% to 60% non condensing
Up to 95% non condensing
STANDARD TEMPERATURE
RANGE
(ALL POWER SOURCES)
OPERATING TEMPERATURE:
TRANSPORT AND STORAGE
TEMPERATURE:
41°F to 95°F (5°C to 35°C)
-2°F to 140°F (-20°C to 60°C)
EXTENDED TEMPERATURE
RANGE:
(USING AC OR DC POWER)
OPERATING TEMPERATURE:
AC SUPPLY:
DC SUPPLY:
95°F to 104° (35°C to 40°C)
Unlimited use all settings, all modes
Unlimited use all settings with pulse mode
Limited to 2.0 LPM, or less, with continuous
mode