Invacare TPO100B Oxygen Equipment User Manual


 
SECTION 8—TROUBLESHOOTING GUIDE
Part No 1156872 41 Invacare SOLO
2
LCD DISPLAY: INDICATORS:
Continuous audible beep every half
second
RED indicator On.
GREEN indicator is Off.
DESCRIPTION: SOLUTIONS:
A stuck button is being detected upon
power up.
1. With the unit turned Off, press each
control button looking for a stuck button.
2. Turn On the unit while ensuring that you
are not continuing to hold the Power
On/Off button once the indicator lights and
LCD begin to illuminate.
3. ONLY press the On/Off button during
startup.
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart. Call your
provider and report the problem if alarm continues.
LCD DISPLAY: INDICATORS:
Triple audible beep every 10 sec
YELLOW indicator Flashing.
DESCRIPTION: SOLUTIONS:
Remaining battery capacity has fallen to
15%. Battery requires charging.
Battery gauge is empty.
Loss of power is imminent.
1. Connect transportable oxygen
concentrator to either AC or DC power,
or use an alternate charged battery
module.
2. If no other power source is available,
Change to another source of oxygen if
alternate power is not available.
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens