MAINTENANCE
Part No. 1145811 35 Stand Up Lift Model RPS350-2
Replacing the Mast Actuator
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
NOTE:Forthisprocedure,refertoFIGURE 7.6.
1. Removethebottomnut,washerandshoulderboltthatsecurethemastactuatortothe
mastmountingbracket.
2. Resttheliftarmonyourshoulderandremovethetopnut,bolt,bracketandbushing
fromtheliftarmmountingbracket.
3. Removethemastactuator.
4. Reversetheabovestepsforinstallationofthenewmastactuator.
FIGURE 7.6 Replacing the Mast Actuator
Bracket
Top Bolt
Lift Arm Mounting Bracket
Top Nut
Bushing
Mast Actuator
Bottom Nut
Washer
Mast Mounting Bracket
Shoulder Bolt