Epson XK3190-H2B Scale User Manual


 
can be settled by increasing division
Err 30 After the datum is stabilized for 30 minutes, the scale body powered off automatically
Err 32 No signal from scale body received, cause by no battery installed or long distance.
Err 33 Accumulation serial No. has reached 99, impossible to accumulate again.
Err 34 Illegal division code, new division need to be entered
Err 35 Accumulation serial No. is 0, deletion is unavailable
Err 36 Accumulation condition unconformable, accumulation unavailable
Err 37 The data entered from keyboard is beyond the range
Err 38 Attempt to modify read-only data
Err 40 The calibrated switch is off, preset data modification unavailable
Err 41 Error password entered
Err 42
Minus ratio:
Caused by (1) Load cell is connected in wrong direction
(2) wrong calibration operation program
(3) The change of inner code is below 5000 after adding standard weight
Err 43 The non-linearity modified value inputted exceeds ±1% Max
Err 50 Zero point exceeds initial zeroing range upon switching on
Err 51 No answer signal received from upper computer during mode transmission
CHAO Weight exceeds Max+9e or preset weight limit
VI Error message from wrong inner operation
Usually, inner operation error is resulted from wrong para.setting, low power, no signal, etc.
Please check and revise the wrong parameter. If cannot be solved, please record all datum and then
contact the Technical Department of Shanghai Yaohua Weighing System Co. Ltd.
Error Message Comments
XXXX 2Error In program address XXXX, exceeding error happens in
subprogram of multiplication
XXXX 3Error In program address XXXX, the error of being divided by zero
happens in subprogram of division
XXXX 4Error In program address XXXX, exceeding error happens in
subprogram of addition and subtraction
XXXX 5Error In program address XXXX, exceeding error happens in
subprogram of BCD code becoming floating code.
XXXX 6Error In program address XXXX, exceeding error happens in
subprogram of floating codes becoming BCD codes
XXXX 7Error In program address XXXX, being powered off automatically 3
minutes later due to low power
XXXX 8Error In program address XXXX, being powered off automatically 3
minutes after no signal received from the scale body.