Energizer 715 Blood Glucose Meter User Manual


 
Example 2: Max Basal
Rusty needs large amounts of insulin to control his blood glucose levels. His new pump was delivered from the
factory with a Maximum Basal Rate of 2.0 units per hour, but he needs 2.8 units per hour in the early morning.
Rusty will reprogram his Maximum Basal to 3.0 units per hour to accommodate his needs.
Stopping your pump
You can stop your pump with the suspend function. Suspend stops all insulin delivery including the current basal
and any bolus or prime deliveries that are in progress. While suspended, your pump will not deliver insulin until you
resume your pump. When you resume your pump, the basal delivery will continue.
The pump will beep or vibrate about every 15 minutes on the hour to remind you that it is not delivering insulin.
Example: You suspend your pump at 11:20 AM. The pump will beep/vibrate at 11:30 AM, 11:45 AM, 12:00 PM, and
so on until you resume your pump (basal resumes).
When suspended, your pump is in Attention mode with a solid circle showing on the pump. When in Suspend, you
can only resume your basal or view the STATUS screen. No other functions are available.
To suspend the pump:
1
Select Suspend from the MAIN MENU, and press ACT.
Main > Suspend
2
SUSPEND will flash on your screen. Press ACT to stop your pump.
3
The screen will show that the pump is suspended and the time that it stopped. After less than one minute, the
pump will return to the HOME screen with a solid circle.
You can verify on your pump STATUS screen that your pump is suspended.
Basic programming 49