Bolus Wizard feature examples
Settings:
• Carb ratio: 30 grams/unit
• BG Target: 90-120 mg/dL (5.0-6.7 mmol/L)
• Insulin Sensitivity: 40 mg/dL/unit (2.2 mmol/L/unit)
• Active Insulin Time: 6 Hours
#1: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter BG.
(food estimate)
60 g
30 g/u
= 2 units
= 2 unitsbolus estimate
#2: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 200 mg/dL (11.1 mmol/
L).
(food estimate)
60 g
30 g/u
= 2 units
= 2 units+
(correction estimate)
= 2 + 2
= 4 units
bolus estimate
11.1 mmol/L - 6.7 mmol/L
2.2 mmol/L/u
= 2 units
200 mg/dL - 120 mg/dL
40 mg/dL/u
Pump specifications 151