• turning off or moving away from other RF transmitting devices; or
• reorienting or relocating the RF glucose meter or pump, or both.
Communications problems may also occur due to cellular telephone interference. Testing conducted with several
different cellular telephones indicates that, while a glucose value is being transmitted, using a cell phone within
12 inches (31 cm) of receiving devices, transmitters or RF glucose meters can interfere with reception of the
transmitted values. In the event of such interference, normal communication can be re-established by:
• turning the cell phone off; or
• keeping the cell phone at least 12 inches away from the pump or glucose meter when a glucose measurement
is being transmitted.
If you have questions, please contact your local Medtronic Diabetes representative.