Add, delete, review remote control IDs
Each remote control has its own unique ID. Up to three different remote control IDs can be programmed in your
pump. The remote control programming screens are very similar to those for the meter. Make sure to select
Remotes when programming your remote control.
If you are not sure that your remote control ID is entered in your pump, check the REVIEW REMOTE ID screen. You
must turn on the remote option to add, delete or review the remote control ID(s) programmed in your pump.
1
Adding a remote ID
a. Select Add ID, then press ACT.
b. Enter each of the six ID numbers found on the back of the remote control. Press ACT after each entry.
After you set the last number of the ID, the screen will return to the REMOTE ID MENU.
2
Deleting a remote ID
a. Select Delete ID, then press ACT.
b. Select the remote ID that you want to delete, then press ACT.
The selected ID is now deleted.
3
Reviewing remote IDs
a. Select Review ID, then press ACT.
b. The programmed IDs will show in the REVIEW REMOTE ID screen.
4
Exit the menus when you are done.
Block feature
Block restricts access to pump programming. The factory setting for this feature is off. Block is an important safety
feature if the pump user requires someone else to maintain complete control of pump operation. When block is on,
the remote control is used to deliver a bolus and suspend/resume the pump. Direct pump programming is limited
to suspend, block, and selftest. You can, however, still view status-type screens (STATUS, BOLUS, BASAL REVIEW,
DAILY TOTALS, etc.). Discuss what uses and settings are best for you with your healthcare professional. (You can
order the remote control from Medtronic Diabetes.)
Turning block on
NOTE:
The vibrate alert-type is disabled when block is on.
1
Go to the BLOCK OPTION screen.