ResMed VPAP III ST-A Sleep Apnea Machine User Manual


 
35System Specifications
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. For further details, see “Guidance and Manufacturer’s
Declaration - Electromagnetic Emissions and Immunity” on page 37.
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6’6” or 9’10” length
IEC 60601-1 Classifications
Class II (double insulation)
Type CF
Continuous operation
This flow generator is not suitable for use in the presence of a flammable
anesthetic mixture with air, or with oxygen or nitrous oxide entrained in the flow
generator airpath.
Table 6: Displayed values
* Results may be inaccurate in the presence of leaks.
For tidal volume up to 800 mL.
For minute ventilation up to 10 L/min.
Note: The manufacturer reserves the right to change these specifications
without notice.
Value Range Accuracy Display Resolution
Pressure sensor at air outlet
Pressure -5 to 30 cm H
2
O ±0.5 cm H
2
O0.1 cm H
2
O
Flow sensor in flow generator*
Leak 0–120 L/min Greater of ±6 L/min
or ±20%
1 L/min
Tidal volume 100–3000 mL ±5%
1 mL
Respiratory rate 6–60 BPM ±0.5 BPM 1 BPM
Minute ventilation 0.6–60 L/min ±7%
0.1 L/min