3Medical Information
Explanation of the warning: When the device is not in operation, and the
oxygen flow is left on, oxygen delivered into the ventilator tubing may
accumulate within the device enclosure. Oxygen accumulated in the device
enclosure will create a risk of fire. This warning applies to most types of flow
generators.
Oxygen supports combustion. Oxygen should not be used while smoking or
in the presence of an open flame.
Always ensure air flow is being generated by the device before the oxygen
supply is turned on.
Always turn the oxygen supply off before stopping the air flow from the
device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on where the oxygen is introduced,
pressure settings, patient breathing pattern, mask selection, and leak rate.
This warning applies to most types of CPAP and bilevel devices.
Do not use the VPAP III ST-A if there are obvious external defects,
unexplained changes in performance, or unusual noises.
Do not open the VPAP III ST-A case. There are no user serviceable parts
inside. Repairs and internal servicing should only be performed by an
authorized service agent.
!
Cautions
You should report unusual chest pain, severe headache, or increased
breathlessness to your physician. An acute upper respiratory tract infection may
require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the
VPAP III ST-A:
drying of the nose, mouth, or throat
bloating
ear or sinus discomfort
eye irritation
skin rashes
chest discomfort.
The above are general warnings and cautions. Further specific warnings,
cautions, and notes appear next to the relevant instructions in the manual.