Philips 862474 Medical Alarms User Manual


 
Performance Procedures
4-24 Testing the Monitor
3 Set the temperature simulator as follows:
a. Temperature: 37
°C (98.6°F).
b. Probe Type: YSI 400 Series
4Press the On/Standby button to turn the monitor On.
5 After the power-up sequence, verify the temperature reads 37°C ±0.1°C (98.6°F
±0.2°F if Fahrenheit is selected as the temperature unit).
6 Turn the monitor Off.
Step Action