Philips 862474 Medical Alarms User Manual


 
Electrical
Specifications
12-3
Electrical
Environmental
Parameter Specification
Power Sources
Internal Battery 12 volt, 4.5 Ah, sealed, lead-acid
Battery Operating Time 3 hours, fully charged battery at
25
o
C(without CO
2
) under normal (one
NBP measurement every 15 minutes, all
other modes active) conditions, 45
minutes if printing continuously, 1 hour,
fully charged battery at 25
o
C (with CO
2
)
under normal conditions, 30 minutes if
printing continuously.
AC Mains 100 to 240 VAC, 50-60Hz, 1A max
Parameter Specification
Mechanical Shock IEC 60068-2-27; 100g, 6 msec, 3 axes, 18
total shocks, non-operating
Mechanical Vibration IEC 60068-2-6; Sinusoidal; 10-58 Hz, 0.15
in displacement; 58-150 Hz, 2g acceleration;
4 min/sweep; 20 sweeps/axis, non-operating
Water Resistance IEC 60529 Classification IPX1 (Protected
against vertically dripping water)
Thermal
Operating 0 to 50
o
C (32 to 122
o
F)
Storage -20 to 60
o
C (-4 to 140
o
F)
Humidity
Operating 5 to 95% RH, non-condensing
Storage 5 to 95% RH, non-condensing