Omega DP119-RTD Thermometer User Manual


 
24
DISPLAY
Type:..........................................................................................7-segment, LED
Height: .......................................................................................0.56" (14.2 mm)
Symbols:..................................................................................................-1.8.8.8
Overrange Indication:....................................Three least-significant digits blank
Colors: .........................................................................................Red - standard
Green - optional
DIGITAL INPUTS
Hold:......................................................................TTL or 5V CMOS compatible
ENVIRONMENTAL
Operating Temperature: ...............................................0¡ to 60¡C(32¡ to 140¡F)
Storage Temperature:.............................................-40¡ to 85¡C (-40¡ to 185¡F)
Relative Humidity: ..............................................95% at 40¡C (non-condensing)
MECHANICAL DIMENSIONS
Bezel: .................................................................... 01.89Ó x 3.78" (48 x 96 mm)
Depth Behind the Bezel: .................................... 4.32" (110 mm) w/ connectors
Panel Cutout: ................................. 1/8 DIN: 01.72" H x 3.62" W (45 x 92 mm)
Weight:............................................................................ 8 ounces (227 grams)
Panel Thickness ........................................................minimum: 0.03" (0.76mm)
maximum: 0.25" (6.25mm)
Case Material: ......................................................94V-1 UL-rated thermoplastic
POWER
AC Frequency: ......................................................................................50/60 Hz
Max. Power: .................................................................................2.4 watts max.
AC Voltage: ................................................................115Vac ±15% 25mA max.
230Vac ±15% 14mA max.
100Vac ±15% 30mA max.
24Vac ±15% 127mA max.
DC Voltage:..........................................................................9-26 Vdc @ 110mA
External Fuse Protection needed:...................................32mA, 115Vac/230Vac
UL Slow-blow (per single unit) 40mA, 100Vac
160mA, 24Vac
125mA, 9-26Vdc
Installation Category: .................................................................. I per EN61010
Equipment Class: .........................................................................I per EN61010
Pollution Degree: ........................................................................2 per EN61010
Refer to the Glossary in Section 6.