Omega DP119-RTD Thermometer User Manual


 
7
3.2 Changing the Meter Parameters
The proper combinations of jumper plugs S1, S2, S3, S4, and S5
inside the meter allow you to change various parameters of the
meter. These parameters include resolution, range, units, RTD
sensor configuration and sensor break configuration. You can
change the parameters by moving the plugs on one or several of
the jumper locations. Keep in mind that the meter is factory set
for 3-wire RTD, and for the upscale position.
Figure 3-2 shows the location of the jumper plugs S1 through S5.
Figure 3-2. Jumper Plugs S1 through S5
P3
P1
W1 W2 W3
2
1
2
3
1
TRANSFORMER
SPAN
DISPLAY
ZERO
(COMPONENT SIDE)
P2
2
3
1
3
4
S2
S2 S3
S1
A
A
AAC
B
B
CD
E
F
GIH
S4 S5
S3
S4
S5
A
C
B
UNUSED PLUG POSITION
S1
ZERO SPAN