Specifications
Result Range: 20 to 600 mg/dL
Calibration: Plasma-equivalent
Sample: Fresh capillary whole blood
Sample Size: Minimum 1 microliter
Test Time: 5 seconds
Assay Method: Glucose oxidase biosensor
Power Source: One replaceable 3.0 V (#2032 or equivalent)
lithium battery
Battery Life: About 1,000 tests (about one year at three tests per day)
Glucose Units: mg/dL
Memory: 150 blood glucose and control solution tests
Automatic Shutoff: Two minutes after last user action
Size: 3.12
X
2.25
X
0.85"
Approximate Weight: 1.5 ounces with battery
Operating Ranges:
Temperature 43–111°F/6–44°C
Relative Humidity 10–90%
Hematocrit 30–55%