Lifescan Ultra Blood Glucose Meter User Manual


 
14
Before you begin:
Use only OneTouch
®
Ultra
®
Control Solution.
Check the expiration date on the control solution vial. Record the
discard date (date opened plus three months) on the vial label.
Do Not use after expiration or discard date, whichever comes first.
Control solution, meter, and test strips should be at room temperature
(68–77˚F/20–25˚C) before testing with control solution (see
Specifications section on page 60 for blood testing temperature ranges).
Shake the vial, discard the first drop of control solution, and wipe off
the tip to ensure a good sample and an accurate result.
Store control solution tightly closed at temperatures below 86˚F
(30˚C). Do Not refrigerate.
WARNING:
Do Not swallow control solution; it is not for human
consumption.
Do Not apply control solution to the skin or eyes as it may
cause irritation.
DCO_06405104B_OTU_Uni_OB_US 3/8/06 12:17 PM Page 14