SECTION 5—WHEELS
Part No 1114851 53 Everyday and Sport Series Wheelchairs
10. Repositionthecamberinsertstothehighestdegreeofcamber.RefertoAdjustingRear
WheelCamberonpage 41.
11. RepeatSTEP3.
12. Closetheleftcamberclamp.RefertoOpening/ClosingCamberClampsonpage 39.
13. Rotatetherighttoeadjustmentringuntilthetabstopsagainsttheuppermetaltabon
thecamberclamp.
14. Securelytightensetscrewonrighttoeadjustmentring.
15. RepeatSTEP7.
16. Performoneofthefollowing:
•ToeIn/ToeOutMeasurementisWithin±1/8inch‐
i. ProceedtoSTEP17.
•ToeIn/ToeOut
MeasurementisNotWithin±1/8inch‐
i. RepeatSTEP1.
ii. Loosenthesetscrewontherighttoeadjustmentring.
iii. RepeatSTEP3.
iv. RepeatSTEPS1216untiltoein/toeoutmeasurementiswithin±1/8inch.
17. Ifdesired,repositioncamberinsertstothedesireddegreeofcamber.Referto
AdjustingRearWheelCamber
onpage 41.
FIGURE 5.17 Adjusting the Axle Tube - A4 Camber System
Camber
Clamp
Higher Degree of Camber
Axle Tube - Will Rotate Down only
Ground/Floor
90° Angle
Place “L” Square Here
Flat Edge of
Camber Tube
Lower Degree of Camber
Right Tab
Stopped
Against
Upper Edge
of Camber
Clamp
Right Toe Adjustment
Ring
Left Tab
Stopped
Against
Lower Edge
of Camber
Clamp
Left Toe
Adjustment
Ring
Camber
Clamp
Axle Tube - Will
Rotate Up only