Fluke ESA612 Thermometer User Manual


 
Electrical Safety Analyzer
Detailed Specifications
63
Ranges................................................................0.0 to 199.9 μA
200 to 1999 μA
2.00 to 10.00 mA
Accuracy
DC to 1 kHz.....................................................±(1 % of reading + (1 μA or 1 LSD, whichever is greater))
1 to 100 kHz ....................................................±(2 % of reading + (1 μA or 1 LSD, whichever is greater))
100 kHz to 1 MHz............................................±(5 % of reading + (1 μA or 1 LSD, whichever is greater))
Mains on applied part test voltage ......................100 % ±5 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI
100 % ±5 % of Mains for 62353 current limited to 3.5 mA ±25 % per 62353
100 % ±5 % of Mains for 60601 current limited to 7.5 mA ±25 % per 60601
Differential leakage
Ranges................................................................10 to 199 μA
200 to 2000 μA
2.00 to 20.00 mA
Accuracy..............................................................±(10 % of reading + (2 counts or 20 μA, whichever is greater))
Insulation resistance
Ranges................................................................0.5 to 20.0 MΩ
20.0 to 100.0 MΩ
Accuracy
20 MΩ Range..................................................±(2 % of reading + 0.2 MΩ)
100 MΩ Range................................................±(7.5 % of reading + 0.2 MΩ)
Source test voltage .............................................500 V dc (+20 %, -0 %) 1.5 ±0.25 mA short-circuit current or 250 V dc selectable
Maximum load capacitance.................................1 μF