Fluke ESA612 Thermometer User Manual


 
ESA612
Users Manual
12
1. From the setup menu, press the softkey labeled
More twice to set F1 to the display contrast function.
2. Press the softkey labeled Display Contrast.
3. Press G or H to increase or decrease the display’s
contrast respectively.
4. Press the softkey labeled Done to exit contrast
setup.
Setting up the Beeper
To enable or disable the beeper:
1. From the setup menu, press the softkey labeled
More twice to set F2 to the beeper on/off function.
2. Press the softkey labeled Beeper to switch the
beeper on and off.
3. Press the softkey labeled Done to go back to the
setup menu.
Viewing Instrument Information
To view information about the Analyzer:
1. From the setup menu, press the softkey labeled
More twice to set F3 to the instrument information
function.
2. Press the softkey labeled Instrument Information.
3. After viewing the displayed information, press the
softkey labeled Done to exit the information screen.
Viewing Memory
Refer to the Using Memory section later in the manual to
learn about viewing memory and how to store data in the
Analyzer.
Setting the GFCI Limit
To set the GFCI current limit:
1. From the setup menu, press the softkey labeled
More to reveal additional menu selections.
2. Press the softkey labeled Instrument to reveal the
instrument setup selections.
3. Press the softkey labeled GFCI Limit to open the
scroll box above the softkey label.
4. Press G or H to adjust the current limit from 5 to
25 mA.
5. Press the softkey labeled Done to exit the GFCI Limit
setup function.
Performing Electrical Safety Tests
The Analyzer is designed to perform a number of different
electrical and performance tests on biomedical
equipment. The following sections describe the various
tests and how to perform them using the Analyzer.