1
Electrical Safety Analyzer
Introduction
The Fluke Biomedical ESA612 Electrical Safety Analyzer
(hereafter the Analyzer) is a full-featured, compact,
portable analyzer, designed to verify the electrical safety
of medical devices. The Analyzer tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The integrated ANSI/AAMI ES1 and
IEC60601-1 patient loads are easily selectable.
The Analyzer performs the following tests:
• Line (Mains) voltage
• Ground Wire (or Protective Earth) Resistance
• Equipment current
• Insulation resistance
• Ground (Earth) leakage
• Chassis (Enclosure) leakage
• Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage
• Lead isolation (Mains on applied parts leakage)
• Differential leakage
• Direct equipment leakage
• Direct applied part leakage
• Alternative equipment leakage
• Alternative applied part patient leakage
• Point to point leakage, voltage, and resistance
• ECG simulation and performance waveforms