Fluke ESA612 Thermometer User Manual


 
ESA612
Users Manual
30
CONDUCTIVE PART
DUT_PE
APPLIED PART
OPEN
TEST LEAD
PE
L2
DUT_L2
DUT_L1
MAINS
L1
EARTH
OPEN
MAINS
ON
L2
OPEN
MD
FE
DEVICE UNDER TEST
APPLIED
PART
faw28.eps
Figure 19. Enclosure Leakage Current Test Schematic
Note
Chassis leakage is the same schematic without the Applied Parts switch.