Basal 67
4. The temporary basal rate will flash. Use the and but-
tons to set the desired basal rate. Press ACT.
Verifying temporary basal delivery
The pump will beep/vibrate three times every hour while deliver-
ing temporary basal rate. You can also verify that a temporary
basal rate is delivering by checking the STATUS screen.
1. From the HOME screen, press ESC to access the STATUS
screen. Use the and buttons to scroll through the
information. Once you have completed your review, press ESC
to return to the HOME screen.
In the example above the Temp Basal is programmed for 0.6 u/h
for 30 minutes. The time left is 24 minutes, this indicates that 6
minutes have elapsed.
MiniMed
SET EDIT TEMP BASAL
03:30
0.
6
U
H
-
MiniMed
STATUS 7:07A U100u
Temp Basal 0.6 U/H
Duration 0:30
Time left 0:24
Last Bolus N 3.8U
9:07A 11 NOV
Basal 1: 0.8 U/H
Units Remaining: 128U
Battery: Normal
Tue 11NOV2000
S/N# 000000
This shows
Temporary
Basal is active
at 0.6 U/H.
Basal rate that
would have
been delivered
if not for the
Temporary
Basal rate.