114 Chapter 9
5. Continue troubleshooting by disconnecting at the
quick-disconnect and set a 10 unit Fixed Prime.
6. Does insulin come out of the needle at the quick-disconnect? If
yes, change your entire infusion set per the instructions in
Chapter 2, Getting Started.
7. If NO insulin comes out of the needle at the quick-disconnect
or you receive another No Delivery alarm call the 24-hour
Product Help-line.
CAUTION: Make sure you reset your Fixed Prime per your infusion set
instructions.
8. Monitor your blood glucose closely.
9. If you have followed these steps and you are still receiving a No
Delivery alarm call the 24-hour Product Help-line.
What happens if I leave the battery out too long?
If you leave the battery out too long (longer than one minute) you
may receive a BATT OUT LIMIT alarm message.
Do the following steps:
1. Set your pump clock to the correct time, date, and year.
2. Check to make sure that all your settings, such as basal rate,
are set as desired.
3. If you do not check your settings you will receive a CHECK
SETTINGS alarm.