30 Chapter 2
Priming with fixed prime
Fixed priming provides a pre-determined volume for priming the
infusion set soft cannula after the introducer needle is removed.
Fixed prime is used after the infusion set has been inserted into
your body and the quick release connected to the pump. A Fixed
Prime is also used if you have disconnected your quick release and
you need to reprime the infusion set.
Refer to your Instructions for Use, for the type of infusion set you
are using, for approximate priming volumes. (Priming volumes
vary depending on type of infusion set you are using.)
The fixed prime cannot exceed the maximum bolus amount you
have set.
1. If two minutes have passed, the HOME screen will appear. (Go
to step 3 if two minutes have not passed.) Press ACT from the
HOME screen, to get back to the PRIME menu. Select Prime
from the Main menu. Press ACT.
2. Select Fixed Prime. Press ACT.
MAIN MENU
Suspend
Basal
MiniMed
Prime
PRIME MENU
Manual Prime
Fixed Prime
MiniMed
Rewind