Energizer MMT-511 Blood Glucose Meter User Manual


 
110 Chapter 8
Deliver an easy bolus
1. With pump in the HOME screen, press and hold ACT to “wake
up” the pump.
2. When the pump is awake, it will beep or vibrate. You will now
be able to program an Easy Bolus.
3. Press the B button for each bolus step desired. Each press will
result in a step beep or vibration. If the beep alert type is
selected, the beeps will change in pitch to assist your counting.
If the wrong button is pressed or the amount is incorrect, simply
wait 30 seconds for the command to time out (do not press any
buttons). To restart programming, you will need to wake up the
pump again.
4. Press ACT. The pump will beep or vibrate back the amount you
just selected. Count the beeps/vibrations. If this amount cor-
rectly equates to the desired bolus amount, press ACT to deliver
the bolus.
Suspend/resume the pump
1. With pump in the HOME screen, press and hold ACT to “wake
up” the pump.
2. To suspend pump delivery, press the S button, then press ACT.
Three audible beeps/vibrations will confirm that the pump is in
Suspend mode.