Energizer MMT-511 Blood Glucose Meter User Manual


 
113
Chapter 9
• • • • • •
Troubleshooting & Care
It is recommended that you read your warranty statement
included with your pump for information on what is covered
during your warranty period.
Troubleshooting
My pump has a no delivery alarm
When a No Delivery alarm occurs, it means the pump is working
correctly. Your pump is not broken, but it has detected that some-
thing is preventing insulin from being delivered.
Do the following steps:
1. Check your blood glucose and take an injection if needed.
2. Make sure that there is insulin in your reservoir and the tub-
ing is not kinked. If these are all right, go to step 5.
3. If necessary unkink tubing. Clear the alarm by pressing ESC
and ACT. A screen will appear with two choices; Resume and
Rewind. Select Resume.
4. If the reservoir is empty, clear the alarm by pressing ESC and
ACT. Select Rewind and change your reservoir and infusion
set per the instructions in Chapter 2, Getting Started.