Barco E-2620 Medical Alarms User Manual


 
Application settings
96 Nio® medical display systems
Application settings
Calibration tab
Figure 49: Calibration tab
Here you can check the Equalization option, allowing you to
match all the displays from a multi-head system to the
luminance of the first calibrated display of the system.
Suppose you work with a dual-head system (2 displays
connected to one display controller card) and you wish
display (2) to have the same output luminance as display (1).
To equalize the displays:
1. Set the Equalization option.
2. Close the Application Settings dialog.
3. Open the Display Settings dialog and calibrate display (2).
The display will be calibrated and additionally the display
luminance will be matched to the luminance of the reference
display (1).