Configuration Commands
C–10
WFM 601A, WFM 601E & WFM 601M User Manual
STANDard <LINES_525 | LINES_625 | AUTO>
Sets the scanning standard that the instrument expects to NTSC (525 lines) or
PAL (625 lines). When in the AUTO mode, the waveform monitor automatically
senses the standard of the input signal. This selection corresponds to the choices
available under STANDARD in the CONFIG FORMAT menu.
STANDard <LINES_525 | LINES_625 | AUTO>
STANDard?
<LINES_525 | LINES_625 | AUTO>
stand auto
stand?
STANDARD AUTO
VECtor:BARS <B100 | B75>
Selects the 75% or 100% graticule for the Lightning and Vector modes. The
graticule must match the type of the selected input signal. This selection
corresponds to the choices available under COLOR BARS in the CONFIG
WFM/VEC menu.
VECtor:BARS <B100 | B75>
VECtor[:BARS]?
<B100 | B75>
vec:bars b75
vec:bars?
:VECTOR:BARS B75
Syntax
Arguments
Examples
Returns
Syntax
Arguments
Examples
Returns