Sunrise Medical P200 Mobility Aid User Manual


 
31
930336 Rev. E
IX. SET-UP, ADJUSTMENT & USE
C. FOLDING BACKREST
Quickie P200 wheelchairs are shipped
with backrest in folded position.
1. Upright Positioning
a) Lift backrest to upright position so
that latch handles (F) lock into place.
b) Backrest is ready for use.
2. Folding Backrest
a) Lift and turn backrest latch handles (F)
inward or outward.
b) Fold backrest forward.
CAUTION: Backrest latches must be
locked into place before operating
wheelchair.
D. FOOTRESTS
1. Installation
a) Place swing-away pivot plate (A) onto
the locating pin (B) on top of the caster
housing with the footrest facing out-
ward from frame.
b) Rotate the footrest inward until
latch plate (C) locks into place on
locking bolt. (D)
Important Note: The swing-away hanger
utilizes a two stage latch.The first stage
locates the locking bolt.The second stage
engages the latch.THE LATCH PLATE MUST
BE SECURELY LOCKED INTO PLACE
BEFORE OPERATING WHEELCHAIR.
2. Removal
a) To remove footrest, push release latch
(C) toward center of chair.
3. Height Adjustment
a) Remove the retaining bolt in the
hanger frame tube using a 7/16"
wrench.
b) Reposition footrest height.
c) Re-tighten bolt.
F
A
B
D
C
D
C
30
930336 Rev. E
IX. SET UP, ADJUSTMENT & USE
A. POWER DRIVE UNIT
1. Wheelchair and power Drive
Attachment
a) Position seat frame over power drive unit.
b) Align quick-connect blocks (A) on
chair with quick-connect sockets (B)
mounted on power drive unit.
c) Rock frame slightly to ensure quick-
connect blocks and sockets are
properly seated.
d) Secure power unit to frame by turning
quick-connect knobs (C) clockwise.
Always secure knobs before operating
chair. If the chair fails to operate and
the battery charge indicator flashes 9
bars, check both quick-connect knobs
for securement.Then cycle the on/off
switch.
B. BATTERY REMOVAL
To remove the batteries for transporting
or servicing:
1. Remove Batteries
a) Unfasten the battery straps (A).
b) Unplug the battery connectors (B) at
the left front of the power unit.
c) Unplug the battery connectors at the
right rear (C) of the power unit.
d) Remove front battery by lifting up on
battery strap (D).
e) Remove rear battery by lifting up on
battery lifting strap (E).
Batteries my weigh up to 55 lbs. (25 kg)
Care must be taken to avoid injury
when lifting.
A
C
B
C
A
D
B
D