English
4466
Weighing and taring
Display view
Function key
The "F" function key is available for controlling functions. It can be assigned various
functions. The functions are changed in the setting menu.
Tip: To additionally display the weight in a second unit, in the setting menu you can assign the F key
the function of activating a second weight display at the top left of the display.
Specify the second weight unit for each measuring point in setting mode under Setting mode/
Scale/Second unit. In addition, the function "Second unit" needs to be activated for the F key:
setting mode: Program Settings → Weighing + Taring → F key → Second unit.
Weighing with manual tare
Determine and tare an unknown bed weight.
Position the empty bed, including bedding, on the scale.
Press the Tare key.
The tare value is saved and the weight display goes to zero.
Now the patient can be laid on the bed.
The weight is displayed with the message "Net".
Clear tare by overwriting with a new tare or by pressing the Zeroing key.
Display the tare memory by pressing the Info key, then press the Tare key. Clear individual points of
the value displayed in the tare memory here using the C key.
The Zeroing key ends the display of the tare memory without change. If you wish to change the
value of the tare memory you can input a new value using the numerical keypad and import it into the
tare memory as a manual tare input using the Tare key. Confirm the changes you have made with the
Print key.