Fluke 1594A Thermometer User Manual


 
83
Remote Operation
Interface Commands
Sets the display Rs Due Alert enable (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default
setting is OFF.
8.7.3.43 DISP:WARN:ITS <boolean>
Query form: DISP:WARN:ITS?
Alternate parameter: DEF
Sets the display ITS-90 Subrange Alert enable (see Section 5.7.4.1, USER SETTINGS (F1), on page 47).
The default setting is ON.
8.7.3.44 FETC? <channel>
Example command: FETCH? 4
Example response: 231.92811,C,4,2009-03-23 14:33:20
Reads a channel’s latest measurement.
The <channel> parameter may be 1 through 4 for corresponding front channel inputs, 5 through 14 for chan-
nels of a rst attached scanner, or 15 through 24 for channels of a second attached scanner. If the <channel>
parameter is omitted, the latest measurement regardless of channel is returned.
If a measurement is available, the response is given in the following format:
<measurement value>,<unit>,<channel number>,<date and time>
<measurement value> is the value of the measurement. It may be a temperature, resistance, or resistance ratio.
The number of decimal places depends on the measured parameter. Trailing zeros are omitted.
<unit> may be “C”, “F”, or “K” for a temperature measurement; “O” for a resistance measurement, or “R” for
a resistance ratio measurement.
<channel number> is 1 through 4 for corresponding front channel inputs, 5 through 14 for channels of a rst
scanner, or 15 through 24 for channels of a second scanner.
<time and date> gives the time and date of the measurement, separated by one space. The time is in 24-hour
format. The date is in the format YYYY-MM-DD.
If no measurement is available, no response is returned and error code -230, “Data corrupt or stale,” is gener-
ated in the error queue. The command STAT:OPER? may be used rst to determine whether a new measure-
ment has completed, and command STAT:QUES:COND? may be used to determine whether the new measure-
ment is valid before an attempt is made to read it.
8.7.3.45 INIT:CONT <boolean>
Query form: INIT:CONT?
Starts or stops measurement. This is equivalent to pressing the START MEASUREMENT or STOP MEA-
SUREMENT function key in the Measurement Screen. A <boolean> parameter value 1 enables measurement;
0 disables measurement.
The *RST command disables measurement.
8.7.3.46 INIT:STOP:BEEP <boolean>
Query form: INIT:STOP:BEEP?
Alternate parameter: DEF
Sets the stop timer’s Audio Indication setting (see Section 5.7.3.2, TIMING SETTINGS (F2), on page 39).
The *RST command sets Timer Beep to the default option which is ON.
8.7.3.47 INIT:STOP:DUR <value>
Example command: INIT:STOP:DUR 1200
Query form: INIT:STOP:DUR?