Fluke 1594A Thermometer User Manual


 
75
Remote Operation
Interface Commands
Bit Event
0 OPC, Operation Complete
1 Request control (not used)
2 Query error
3 Device dependent error
4 Execution error
5 Command error
6 User request (not used)
7 Power on
8.7.3.4 *IDN?
Example response: FLUKE,1594,A9C111,1.10/1.00
Reads the instrument information.
The response is given in the following format:
<manufacturer>,<model number>,<serial number>,<version numbers>
8.7.3.5 *RST
Sets operating settings to default states.
The following settings are affected:
Any self-test in progress is canceled
Measurement enable is set to OFF
Channel 2 enable is set to ON, all other channels OFF
Scan Enable is set to ON
Repeat Count is set to 1
Alternate Reference is set to OFF
Reference Channel is set to channel 2
Timing settings are preset to NORMAL presets
Stop Timer Enable is set to OFF
Timer Duration is set to 3600 s
Timer Beep is set to ON
Filter Enable is set to ON
Time Window is set to 60
Statistics Mode is set to UNLIMITED
Sample Count is set to 100
Delta Mode is set to OFFSET
Delta Offset is set to 0.
Delta Channel is set to channel 2
Unit is set to °C
The key lock setting is set to off
Statistics are reset
Not affected are channel settings other than enable, probe denitions, resistor denitions, recording settings,
display settings other than Unit, remote interface settings, time and date, password settings, and calibration
settings.
8.7.3.6 *SRE <value>
Example command: *SRE 16