ETA Systems FC01-LQ Oxygen Equipment User Manual


 
Flow Meter FC01-LQ Flow Meter FC01-LQ
56 Technical data 56 Technical data
8.6 Sensor interface
8.6.1 Electrical data of the terminal for calorimetric monitoring heads
Terminal Mnemonics Data
XSK1 R(HEIZ)-LO Function: terminal for negative pole of heater element
Drain output of heating current control
Max. sink current: I
sink
= 88 mA
Dielectric strength: -0.5 V … +20 V DC
XSK2 R(HEIZ)-HI Function: terminal for positive pole of heater element
Hi-potential of heater source
Output voltage range (load dependent)
U
a
= 21 V … 24 V DC
Max. current output: I
max
= 100 mA
Not short-circuit proof
XSK3 R(Tref)-HI Function: terminal for positive RTD * pole for medium
temperature measurement
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK4 R(Tref)-LO Function: terminal for negative RTD * pole for medium
temperature measurement
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK5 AGND Function: analogue ground
Reference potential of exitation current source for
RTD * operation
XSK6 IS Function: output of exitation current source for
RTD * operation
Exitation current: 1 mA ± 1%
Admissible load range: R
load
= 0 … 2 k
Dielectric strength: ±15 V DC
XSK7 SGND Function: shield ground
XSK8 Terminals for sensor cable shielding
XSK9 R(Tdiff)-LO Function: terminal for negative pole of the heated RTD *
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK10 R(Tdiff)-HI Function: terminal for positive pole of the heated RTD *
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
* RTD = Resistive Temperature Device
8.6 Sensor interface
8.6.1 Electrical data of the terminal for calorimetric monitoring heads
Terminal Mnemonics Data
XSK1 R(HEIZ)-LO Function: terminal for negative pole of heater element
Drain output of heating current control
Max. sink current: I
sink
= 88 mA
Dielectric strength: -0.5 V … +20 V DC
XSK2 R(HEIZ)-HI Function: terminal for positive pole of heater element
Hi-potential of heater source
Output voltage range (load dependent)
U
a
= 21 V … 24 V DC
Max. current output: I
max
= 100 mA
Not short-circuit proof
XSK3 R(Tref)-HI Function: terminal for positive RTD * pole for medium
temperature measurement
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK4 R(Tref)-LO Function: terminal for negative RTD * pole for medium
temperature measurement
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK5 AGND Function: analogue ground
Reference potential of exitation current source for
RTD * operation
XSK6 IS Function: output of exitation current source for
RTD * operation
Exitation current: 1 mA ± 1%
Admissible load range: R
load
= 0 … 2 k
Dielectric strength: ±15 V DC
XSK7 SGND Function: shield ground
XSK8 Terminals for sensor cable shielding
XSK9 R(Tdiff)-LO Function: terminal for negative pole of the heated RTD *
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
XSK10 R(Tdiff)-HI Function: terminal for positive pole of the heated RTD *
Input resistance: > 1 G
Dielectric strength: -17 V … +30 V DC
* RTD = Resistive Temperature Device