Braun 950787 Thermometer User Manual


 
WARNINGS & CAUTIONS
Vista
basic Instructions for Use 4/25
950787 Rev H 02/03
After starting an infusion, make sure
drops are falling in the drip chamber.
If no drops are falling, make sure the
roller clamp is open. If the roller
clamp is open and still no drops are
falling, replace and dispose of the
set.
The air detector cannot recognize
the introduction of air at 3-way
stopcocks, infusion ports, and other
lines/tubes below the pump.
See section entitled “Drop Sensor”
for additional information on using
the device without a drop sensor.
To avoid mechanical or electronic
damage, do not steam autoclave or
immerse the pump in any fluids or
cleaning solutions. Always
disconnect electrical power cord
from outlet before cleaning to
prevent electrical shock.
CAUTIONS
Rx only.
Do not attempt to infuse two fluids
simultaneously using the Vista
basic.
Variations in pressure, e.g. caused
by change in level of fluid container,
may affect the accuracy of the
device.
If several pieces of infusion
equipment are connected together,
it is possible there may be mutual
influence. Possible incompatibilities
can be found in the instructions for
use for the drug or for the other
appliances.
Use only equipment, accessories,
parts, and disposables that are
intended for use with the Vista
basic.
Do not clean, disinfect, or sterilize
any part of the device by autoclaving
or with ethylene oxide gas. Doing
so may damage the device and void
the warranty. Only external parts of
the pump should be disinfected.
See “Maintenance, Storage, &
Service” for suggested cleaning
solutions.
Technical specifications are not
guaranteed with any other tubing
than that approved for use with the
Vista basic.
Any analog and digital components
connected must demonstrate that
they meet EN specifications (i.e. EN
60950 for data processing
equipment and EN 60601 for
electrical medical equipment).
Only an authorized system
configuration technician responsible
for meeting the specifications of
standard EN 60601-1-1 may
connect additional devices to the
signal inputs or outputs.
SAFETY STANDARD
The Vista basic meets all safety
standards for medical electrical devices,
corresponding to IEC 60601-1 and IEC
60601-2-4.
NOTE: IEC 60601-1 corresponds to
European Standard EN 60601.