BOLUS MODE
This feature allows a bolus to be given during a Continuous Mode infusion.
1. Press OPTS until display shows BOLUS MODE.
2. Press ON.
3. Press RATE.
4. Enter the rate at which the bolus should infuse, then press OK.
5. Press END.
6. To give a bolus during the infusion, press BOL.
7. Enter the amount of bolus to be given in ml.
8. Press OK.
Bolus Without Entering a Volume
1. During the infusion, press BOL and hold down until a second BOL is
displayed to the right.
2. Continue to hold down the first BOL while also pressing the second BOL.
A bolus will be given as long as both BOL keys are held down. The pump
will display the amount of bolus being administered and will beep once after
each milliliter delivered.
INTERVAL BOLUS (INTERMITTENT MODE)
The basal/KVO rate and entire bag volume must be entered in CONTINUOUS
MODE first.
1. Press OPTS until display shows BOLUS MODE.
2. Press ON.
3. Press TIME.
4. Enter the time between the start of each dose, then press OK.
5. Press RATE.
6. Enter the rate at which the dose is to be given, then press OK.
7. Press VTBD.
8. Enter the volume of each dose, then press OK.
9. Press END.
10. Press . The pump will deliver at the basal/KVO rate for the interval
time, then begin delivery of the first dose.