AC and 12 V
Check AC voltage as per
rating label.
Insert/screw on power lead
(12 V lead in ambulance car)
on back.
Battery Operation
Switching to internal battery is
automatic in the event of
power failure or if 12 V or AC
is not connected.
Charge battery on initial use,
following a battery alarm, or
after 2 months of non-use.
Charging time: approx. 16
hours. Longer charging has
no detrimental effect.
Full battery provides operation
for 3.5 hours at 125 mL/hr.
Battery capacity is approx.
50% after 2 years.
The service life of the battery
can be extended by
completely discharging it from
time to time and then fully
recharging it on AC power.
CC Interface
Interface descriptions available from
B. Braun. Connection to interface
input uses MFC plug.
Documentation
All operating data of the
Vista
basic can be called up
and logged via external
computer