Braun 950787 Thermometer User Manual


 
OPTIONS Total Infused, Standby Function
Vista
basic Instructions for Use 18/25
950787 Rev H 02/03
TOTAL INFUSED
The Total Infused information screen displays the total amount of fluid in ml that
has been pumped since the last time the Total Infused was cleared.
To Clear the Total Infused
1. Press OPTS until display shows Tot Infused=.
2. Press .
3. Press END.
STANDBY
This function allows the pump to be put on standby while retaining all set infusion
data. This function is only available in Continuous Mode.
1. Press OPTS until display shows STANDBY
2. Press ON.
3. Enter time in hours and minutes the pump is to stand by.
4. Press C to clear the current data and re-enter new data. Press OK to
accept new data.
4. Press OK.
The pump will count down to zero and then alarm Pump Stopped!
To Exit Standby Function
1. Press EXIT to return to Continuous Mode.
Note: If countdown has already started, press END to return to Continuous
Mode.