Mechanism
Linear peristaltic volumetric infusion pump
Dimensions WxHxD
5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm)
Weight
6.8 lbs. (3.1 kg)
Power Requirements
100-120 VAC, 50-60 Hz, 13 Watts
Leakage Current
Max 30µA
Battery
7.2V, 1.2 Ah NiCd (rechargeable)
Battery Operating Time
=3.5 hours at 125 mL/hr
Rate Range
0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™
tubing
Volume to be Infused
Range
0.1 to 9999.9 mL
Occlusion Pressure
Variable Settings Low: 8 PSI High: 17 PSI
KVO Rate
1 mL/hr or selected rate, whichever is lower
Flow Rate Accuracy
± 5% measurement time 8hr, temp. 71.6° F
Alarms
Air-in-Line, Low Battery, Plug In Now, Bag Empty,
Downstream Occlusion, Door Open, Pump Stopped,
KVO, Unit Malfunction
Air-in-Line Alarm
Air bubble size 0.3mL* or 1.5mL/hr**
*Can be set from 0.01 to 0.3 mL in service
program
**Can be set from 0.5 to 3.5 mL/hr in service
program
Memory
User option to retain last Continuous and Piggyback
Mode data upon power up. Permanent data
retention in Program and Ramp/Taper Modes after
pump is turned off
Tubing
B. Braun Vista™ administration sets
Modes/Features
Continuous and Piggyback Modes, Program Mode
for variable infusions, Ramp/Taper Mode for TPN
infusions and Delay A Mode for postponing infusion
start and finish
Operating
Temperatures
Operating: 64.4° F to 95° F, 10.15 PSI to 15.37PSI
Non-Operating: -13° F to 131° F, 10.15 PSI to
15.37 PSI