Bayer HealthCare Blood Glucose Meter MODEL Ascensia Elite Blood Glucose Meter User Manual


 
9
The Control Test SECTION 2
To make sure your Ascensia ELITE™ Diabetes Care System is performing properly, a Control
Test is performed using the Ascensia E
LITE™ Normal Control solution.
WHY? A Control Test confirms that the system is functioning properly and you are using
proper technique. When running a Control Test, the system is defined as: the Meter and
the Test Strips. In addition to the Normal Control, an Ascensia E
LITE Low and High Control
are also available for use with this system.
WHEN TO RUN A CONTROL TEST:
• Any time you want to confirm that the Meter system is working properly, and that your
testing technique is correct.
• Any time you question your blood glucose test results.
• On the advice of your physician or healthcare professional.
WHAT TO EXPECT:
• When the Control Test is done properly, the Control Test result should fall within the
range listed on the end flap of your Test Strip Carton.
• A Control Test result that falls outside of the specified range means the system may not be
functioning properly or your testing technique may not be correct. If this occurs, refer to
Section 6 of this User Guide or the Ascensia E
LITE Control package insert for assistance.
MARK THE CONTROL TEST RESULT (if using the memory feature):
You should always “mark” a Control Test result with a “C” as described in the following
procedure. Marking with a “C” keeps a Control Test result from being included in the
average of your blood glucose tests.