OVER
Steps Do This
5. Set, review, or change your
Projected Low glucose alarm.
a. Select Projected Low from the Alarms menu.
•UsetheUP/DOWNArrowbuttonstohighlight
Projected Low.
•PresstheRIGHT Option button to select
Projected Low.
b. Review or change your settings.
•PresstheLEFT Option button (Next) to move
from one eld to the next.
•UsetheUP/DOWNArrowbuttonstochange
the setting in any eld.
•PresstheRIGHT Option button (Set) to accept the
settings and return to the Alarms menu.
Note: e receiver will respond with a brief sample
of the alarm you selected.
c. Decide what to do next.
•Youaresetting,reviewing,orchanging
all alarms.
Continue to the next step.
•Youarenishedsetting,reviewing,and
changing alarms.
Go to Step 11.
If en
Decide what settings you want:
• Alarmonoro.
• Sensitivity.
• Alarmtype(Setting).
– Low, medium, or high beep.
– Short, medium, or long
vibration.
If you want this much notice
before reaching your Low
glucose level:
en set the
Sensitivity to:
•Maximumdetection(most
alarms, approximately 30
minutes in advance of events)
•Intermediatedetection
(intermediate alarms,
approximately 20 minutes in
advance of events)
•Lowdetection(fewestalarms,
approximately 10 minutes in
advance of events)
High
Medium
Low
Section 8 73
Set, Review, or Change the Alarm Settings
Alarm Settings 08:30A
Low Glucose (70)
High Glucose (275)
Projected Low (On)
Projected High (O)
Data Loss
Main Select
Alarm Settings 08:30A
Projected Low Glucose
ON
Sensitivity: Low
Setting: Medium Beep
Next Set