Figure 2-3 YSI 600 Sonde Calibration Flow Chart
Program Step #3, “A djust Level/Parameters”
Parameter To Adjust
None - Level-YSI 600 - Ph - D.o.
YS I 600 Parameter To Calibrate
None - Ph - Conductivity - D.o.
YS I 600 Ph Calibration
pH4 & pH7 - pH7 & pH 10 - pH4, pH7, & pH10
Rinse probe and place in ph 4.0 solution
Press Enter when stable: X.XX pH
Calibrating...
Please Wait...
Error/warning Override
Yes - No
Rinse probe and place in ph 7.0 solution
Press Enter when ready: X. XX pH
Calibrating...
Please Wait...
Er ror/warning Overri de
Ye s - No
Rinse Probe And Place In Ph 7.0 Solution
Press Enter when stable: X. XX pH
Calibrating...
Please Wait...
Error/warning Override
Ye s - No
Ri nse probe and place in ph 10.0 solution
Press Enter when stable: X.XX pH
Error/warning Override
Ye s - No
Rinse probe and place in ph 4.0 solution
Press Enter when stable: X.XX pH
Er ror/warning override
Ye s - No
Rinse probe and place in ph 7.0 solution
Press Enter when stable: X.XX pH
E rror/warning override
Ye s - No
Ri nse probe and place in ph 10.0 solution
Press Enter when stable: X.XX pH
Error/warning override
Ye s - No
No
Ye s
No
Ye s
Good Ca librat ion
E rror/
Wa rni n g
Error /
Wa r n i ng
Er ror/
Wa rnin g
Error/
Wa r n i ng
Error/
Wa r n i ng
Error /
Wa r n ing
E rror/
Wa rnin g
Good Cali brat ion
Good Cali brat ion
Good Cal ibrat ion
Good Cal ibrat ion
Good Cali brat ion
Good Cal ibrat ion
No
No
No
No
No
Ye s
Ye s
Ye s
Ye s
Ye s
Calibrating...
Please Wait...
Calibrating...
Please wait...
Calibrating...
Please wait...
Calibrating...
Please wait...
YS I 600 Parameter To Calibrate
None - Ph - Conductivity - D.o.
YS I 600 D.O. Calibration
D.O. Standard - Abs. Barometric Pressure - Altitude
Place Probe in X.XX Mg/l solution
Press Enter when ready:
Error/warning Override
Ye s - No
Units for altitude entry
Ft. - M
Error/warning override
Ye s - No
Enter conductivity calibration units
mS/cm - ppt
Error/warning Override
Ye s - No
No
Ye s
No
Ye s
Good Cali brat ion
Error/
Wa r n i ng
Er ror/
Wa rni n g
Error/
Wa r n i ng
Good Cali brat ion
Good Calib ration
No
Ye s
D.O. Standard
X.XX Mg/l
Abs. Barometric Pressure
X.XX mmHg
Place probe in calibration cup
Press Enter when ready:
Error/warning Override
Ye s - No
No
Ye sGood Calib ration
Error/
Wa r n i ng
Place probe in calibration cup
Press Enter when ready:
Enter conductivity standard
XXX Units
Place probe in XXX units solution
Press Enter when ready
Calibrating...
Press Enter when stable X.XX Mg/l
Calibrating...
Press Enter when stable X.XX Mg/l
Calibrating...
Press Enter when stable X.XX
mS/cm
Calibrating...
Press Enter when stable X.XX Mg/l