11. Determine the percent error by using the formula below:
%Error + [
M1 * M2
0.4
* 1] 100%
The calculated error should be 2%.
12. Record the results on the test record.
NOTE. An unaccceptable error value may result if a low tolerance
termination is substituted for the recommended termination.