P6249 4 GHz 5X Active Probe Instruction Manual
To set the probe offset, follow these steps:
1. Set the oscilloscope coupling to GND.
2. Use the vertical position control to set a zero reference level on
the oscilloscope display.
3. Set the oscilloscope coupling to DC and 5 V/div.
4. Attach the probe to the circuit.
5. Adjust the probe offset to bring the trace to the oscilloscope zero
reference.
6. Change the volts/division setting to the desired range, adjusting
the offset to keep the trace on the zero reference level.
NOTE. The P6249 has a ±5.0 V offset range. The linear operating
range is ±2.0 V. See Figure 1. Also, see page 19 for more informa-
tion.
If cursors are used on a TEKPROBE oscilloscope, the zero reference
will be at the probe offset voltage.
0 V
+7.0 V
–7.0 V
+2.0 V
–2.0 V
+5.0 V
–5.0 V
Nonoperating range (+30 V maximum nondestructive input voltage )
+5.0 V
–5.0 V
Nonoperating range (–30 V maximum nondestructive input voltage )
Maximum offset rangeMaximum AC signal amplitude
Figure 1: Dynamic and offset limitations