ResMed 338425/1 Sleep Apnea Machine User Manual


 
32System Specifications
Noise Level
Accessories
Data Card
Dimensions (L x W x D): 65.04 mm x 22.15 mm x 0.85 mm
Data Card Module
Dimensions (H x W x D): 49.7 mm x 42.2 mm x 12 mm
All Devices (where applicable)
Housing Construction: Flame retardant engineering thermoplastic
Environmental Conditions
Operating Temperature: +5°C to +36°C
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -20°C to +60°C
Storage and Transport Humidity: 10%–95% non-condensing
Operating Altitude: sea level to 2591 m
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. For further details, see “Guidance and Manufacturer’s
Declaration — Electromagnetic Emissions and Immunity” on page 34.
Note: The manufacturer reserves the right to change these specifications
without notice.
AutoSet Spirit II—CPAP mode, 10 cm H
2
O set pressure
*
* Declared dual number noise emission values in accordance with ISO 4871.
Measured total A-weighted sound pressure level
(dB re 20 μPa)
24 dBA
Uncertainty (dB re 20 μPa) 2 dBA
Measured values are determined according to the noise test procedure given in
ISO 17510-1.
Note: The sum of a measured noise emission value and its associated
uncertainty represents an upper boundary of the range of values that is
likely to occur in measurements.