11.6TraceabilityofInstrumentCalibration
CUSTOMER INFRARED THERMOMETERS
Computerized cavity
(black body source)
500 to 2727
°
C
400 to 800°C
Medium Temp Secondary
(Working) Standard
400 to 800°C
Fluke Medium Temp
Primary Standard
760 to 2700°C
High Temp
Primar
y
Standard
760 to 2700°C
High Temp Secondary
(Working) Standard
Working cavaties and plates
(black body sources)
25 to 500°C
-20 to 800°C
Low Temp Secondary
(Working) Standard
-20 to 800°C
Low Temp
Primary Standard
Daily check at 1000,
1700, and 2500°C
Cross check
Annual NIST Calibration
and historical data
results review
Annual NIST Calibration
and historical data
results review
Fluke Calibrated
blackbody cavaties
and
p
lates
Fluke Working
Standards
Fluke Primary
Standards
National Metrology
Institutes
PTB
-50 to 800°C
ITS 90
NIST
800 to 2700°C
ITS 90
Fluke reserves the right to make changes to this traceability and methodes without notice.
Fluke IR Traceability